I believe that many customers will encounter the phenomenon of undesired hop counts when they use electronic scales. In response to this phenomenon, today we are fortunate enough to have invited our professional engineers from Shanghai Atsujin Weighing Equipment Co. Several reasons for the number:
1 The digital strobe is generally the voltage of the instrument is not enough, caused by the low voltage, and the instrument's built-in battery and charger are normal.
2 If the instrument does not move randomly, check whether the signal line of the instrument is disconnected and damaged, and whether there is water in the junction box and a short circuit.
3 Electrostatic interference on the scale body. Check the grounding wire and grounding.
4 sensor water damage, need to replace the sensor.
5 The number of the instrument has a significant increase or decrease during the weighing process. Generally, the sensor is unhealthy and the sensor must be replaced.
6 The instrument AD is bad, then it needs a professional to overhaul it. Say so much, I believe that you already have a basic understanding of the reason for the number of hops. If you encounter any problems in the future use, please call us. Tianjin Weighing Instrument Free Advice